Meet Standard
ASTM G155: Standard Practice for Operating Xenon Arc Light Apparatus for Exposure of Non-Metallic Materials.
ASTM 04459: Standard Practice for Xenon-Arc exposure of plastics intended for indoor applications.
ISO 4892-1 Plastics - Methods of exposure to laboratory light sources.
ISO 16474-2 Paints and varnishes - Methods of exposure to laboratory light sources - Part 2
Model | XLF - Volume | |
Internal Dimension WxHxD (mm) | Customers designated space size and design | |
Tube Using Life | 1200H | |
Tube Power | 1KW / 6KW | |
Illumination Intensity | 340nm 0.35 ~ 0.6W/m² |
Full Spectrum 1200±200w/m² |
Lighting Filter Type | Inner Filter Outer Filter |
Quartz |
Simulated Test Status | Test samples explore under outer space | |
Cooling System | Wind cooling or water cooling | |
Temperature Control Range of the Test Cabinet |
Black panel temperature : BPT 40~120℃(±3°C) | |
Black standard temperature : BST 38~120℃(±3°C) | ||
Humidity Control Range of the Test Cabinet |
Humidity when lighting : 20%R.H. ~ 75%R.H. ( ± 5%R.H.) | |
Humidity when dark : 60%R.H. ~ 95%R.H. ( ± 5%R.H.) | ||
Rotating Speed of the Samples Holder |
1~5 RPM ; adjustable | |
Diameter of Sample Shelf | 500±50mm | |
Sunlight Exposure Cycle | 0.01~ 999 hours 59mins | |
Raining Period/ Time | Raining Period: 1~240mins,discontinuous-type, adjustable | |
Raining Time : 1-9999M, adjustable , water spray pressure70~130Kpa | ||
Spray Water Flow | 1L ~ 2.4L/min | |
Explore Area | 7500cm² | |
Power | AC 380V: 3∮ 5 lines : 50/60HZ |