Test EQ

IEC60068-2-66 Pressure Accelerated Aging Test Chamber (HAST)
Release time:  2025-06-03 10:56:35

Pressure Accelleration aging test chamber -HAST Environmental testing Part 2: Test methods Test Cx: Damp heat, steady state (unsaturated pressurized vapour) PCT Chamber, HAST chamber

TEST EQ Pressure accelerated aging test chamber(HAST chamber) strictly follows IEC 68-2-66 international standard, simulates high temperature, high humidity and high pressure environment, accelerates product life test. The Test chamber is equipped with precise temperature and humidity control (105°C-150°C, 75%-100%RH), rapid voltage rise (≤15 minutes), and multiple safety protection systems, making it suitable for reliability verification of semiconductors, PCBS, and electronic components. Through HAST/PCT test chamber , the customer can reduce the traditional 1000-hour aging cycle to 96 hours, significantly improving R&D efficiency and reducing quality inspection costs

Pressure Accelleration aging test chamber -HAST

1 Scope 

This International Standard provides a standard test procedure for the purpose of evaluating,  in an accelerated manner, the resistance of small electrotechnical products,  primarily nonhermetically sealed components,  to the deteriorative effect of damp heat. 

The test is not intended to evaluate external effects,  such as corrosion and deformation.


 2 General description 

In this test,  the specimen is subjected to very high levels of unsaturated damp heat for a relatively short period. 

Electrical bias is  usually applied. 

Due to the very accelerated nature of the test,  careful consideration must be given to the choice of test conditions,  since this may have an important effect on the type of failure mode which could occur (see annex B). 

The test provides  three temperatures at a relative humidity of 85 %. The test severity is defined by the duration at one of the  temperatures. 

Care shall be taken not to reach the maximum rated temperature of the specimen and/or the critical  temperature of any encapsulating materials. The glass transition temperature of a plastic is a typical example of a  critical transition temperature. 

In the case of plastic encapsulated components,  degradation results from absorption of water-vapour by the plastic,  and penetration of moisture along terminals. 

3 Description of test apparatus 

3.1 The test chamber The chamber shall be so constructed that:  

a) it can produce the temperature and relative humidity given in table 1,  and sustain the pressure values given in clause 4, note 3;

b) it is capable of providing controlled conditions of temperature, relative humidity and pressure during testing,  and the ramp-up to, and ramp-down from, specified test conditions; 

 c) the temperature and humidity of the chamber can be monitored by means of sensing devices located in the working  space,  and/or other areas giving the same results (e.g. a steam generator). 


NOTE-The direct measurement of relative humidity  during this test is not possible with current techniques. Guidance on the determination of the relative humidity in the  working space is given in annex C.


d) the enclosed atmosphere can be expelled from the chamber by the water-vapour prior to the commencement of testing;  

e) condensed water is not allowed to fall on the specimen;  

f) the materials used in the construction shall not cause any corrosion of the specimen,  or degradation of the quality of the humidifying water (see clause D.2). 


The temperature tolerance of t2 C is intended  to take account of absolute errors in the measurement, fluctuations of the chamber temperature at any point,  and variations between all points within the working space. 

However,  in order to maintain the relative humidity within the specified tolerance of t5 %,  it is necessary to keep the difference between any two points in the working space (at any instant) within narrower  limits. 

The specified humidity tolerance will be exceeded if such temperature differences exceed 1.5 °C. It is also necessary to restrict the short-term temperature fluctuations due to cycling of the chamber heater to a similar value.  

The specimen should not significantly impede the vapour flow. 

Condensation shall not be allowed to form on the specimen  at any time during the test. 

3.2 The humidifying water 

Distilled or deionised water shall be used. The water shall have a resistivity of not less than 0.5 MΩcm at 23 °C. The pH value shall be between 6,0 and 7,2 at 23 °C. 

Before the water is placed in the humidifier,  all internal parts of the chamber shall be cleaned. Guidance on cleaning is given in clause D.4.


4 Severities 

One of the combinations of temperature and duration shown in table 1 shall be used unless otherwise  specified in the relevant specification. For each temperature, three durations are defined.

IEC60068-2-66-TESTEQ-HAST-CHAMBER.jpg


 NOTES د

1  It is not recommended that a test should be restarted; however, if,  at a given temperature it is required to subject the specimen to a longer duration than that given in column Ill of  table 1,  then the test shall be recommenced in accordance with the requirements of clause 6. The test shall be recommenced within  96 h of the end of the ramp-down period of the previous test. 

During the interval between the tests,  the specimen shall be held under standard atmospheric conditions for measurement and tests,  unless otherwise specified in the relevant specification. 

2 The durations given for testing at the specified conditions  do not include the additional time required for the ramp-up and ramp-down period and the chamber cleaning and  preparation time 

3 The vapour pressures are approximately 0.12 MPa, 0.17 MPa and 0.23 MPa at 110 C 120 C and 130 °C respectively.


6 Testing 

6.1 With the chamber and specimen at the laboratory conditions of temperature, pressure and humidity the specimen shall be installed in the working space in the chamber. 

6.2 Positioning and mounting of the specimen The specimen shall not be subjected to radiant heat from the heaters or the chamber wall. 

If required by the relevant  specification,  a specific mounting structure shall be used. The heat conductivity and the thermal capacity of the mounting fixtures  shall be low so that for all practical purposes the specimen is thermally isolated. 

Care shall be exercised in the  choice of mounting structure and mounting fixture materials to minimize the effects of contamination and to minimize  degradation due to corrosion and other mechanisms (see clause D.2). 

6.3 Bias voltage 

If required by the relevant specification,  the specimen shall have a bias voltage applied during exposure. Guidance on the application of bias is given in annex D.  

The bias voltage (or bias voltage cycle) shall be applied to the specimen when the temperature and the relative  humidity have reached the stable state, and continue until the specimen is under recovery conditions.

6.4 The test cycle 

6.4.1 The temperature of the chamber shall be raised to the appropriate value During this period  enclosed atmosphere is expelled from the chamber by the water-vapour,  and the temperature and the relative humidity shall not exceed the specified value. Condensed water is not allowed on  the specimen at any time during the test cycle. Stabilization of temperature and humidity shall take place within 1,5 hours. 

However, if stabilization takes longer than 1,5 hours the test requirements will be met if stabilization is achieved in not more than 3 h and the specified test duration is  greater than 48 h. 


6.4.2 The temperature and relative humidity shall be maintained within the prescribed limits for a duration as specified in the relevant specification. The test duration shall commence as soon as the conditions have  stabilized. 


6.4.3 At the end of the specified duration, the chamber pressure temperature and relative humidity shall be restored to standard atmospheric conditions for measurement and tests in not  less than 1 h and not more than 4 h. 

During this period,  the temperature and the relative humidity shall not exceed the specified value. The conditions may be achieved by means  of natural cooling. 

If the pressure is reduced by release from the chamber,  care shall be taken that the specimen is not subject to rapid depressurization. The pressure shall not fall below the  laboratory pressure. The bias should normally be maintained during this period. 


6.4.4 On completion of the cooling period,  the specimen shall be subjected to the recovery procedure. 


7 Intermediate measurements 

The relevant specification may  require electrical and/or mechanical checks during testing. 


If it is required to make intermediate measurements,  the relevant specification shall define the measurements and period(s) during testing after which they shall be carried  out. The measurements shall not cause any change to the test conditions. 

Measurements preceded by recovery which would  require the removal of the specimen from the chamber are not permissible during testing.

IEC-68-2-66-TESTEQ-Pressure-acceleration-aging-chamber-HAST-Chamber3.jpg

IEC-68-2-66-Steam-PCT-HAST-chamber 1.jpg

IEC-68-2-66-TESTEQ-HAST-PCT-Chamber-2.jpg

Act now to enhance product reliability! Order HAST/PCT test chambers this month and enjoy free standard calibration service plus warranty on core components! 

Submit inquiry for exclusive solutions:  Click to fill in the requirements form | Email : kris@testeq.com  

⚠️ The first 10 inquirers of today will receive an additional IEC 68-2-66 test guidebook!


  • Add: TESTEQ Science and Technology Park, No. 5, Wenzhou Road, Dongcheng Sub-district, Dongguan City, Guangdong Province, China

    Whatsapp(Phone): +8613310826672

    Tel: +86-0769-85914911

    Web: www.chamber-testing.com

    E-mail: Kris@test-eq.com

Ask For A Quick Quote

We will contact you within 1 working day, please pay attention to the email with
the suffix "@test-eq.com".

We will contact you within 1 working day, please pay attention to the email with the suffix "@test-eq.com".

Company
Name*
Phone
E-Mail*
Nation
Message
By selecting the ‘I agree’ button, you consent to the processing of your personal data for marketing purposes and for the sending of commercial information by TESTEQ Company through telephone, e-mail, newsletters, text messages such as sms, chat and social networks.
I AGREE
By submitting your personal data, you declare that you have read the Privacy Policy of the website.

P.IVA: 91441900MAECQJHY37 Guangdong TESTEQ Testing Equipment Co., LTD | All Rights Reserved